A suture anchor
HomeHome > Blog > A suture anchor

A suture anchor

Jan 24, 2024

Scientific Reports volume 13, Article number: 3493 (2023) Cite this article

481 Accesses

Metrics details

The aim of this biomechanical investigation was to evaluate a repair technique for type IV FDP tendon avulsions using a suture anchor, addressing the bony and the tendinous aspect of this injury simultaneously. In 45 distal phalanges from human anatomical specimens the injury was simulated and repairs were performed with a suture anchor using an innovative technique, interosseous sutures and a combination of screws and an interosseous suture. Repetitive loading for 500 cycles simulated postoperative mobilization. Repairs were loaded to failure thereafter. Elongation of the tendon-suture complex, gap formation at the bone-bone contact line and at the bone-tendon insertion line, load at first noteworthy displacement (2 mm), load at failure and the mechanism of failure were assessed. The suture anchor technique was superior biomechanically considering load at failure (mean: 72.8 N), bony gap formation (mean: 0.1 mm) as well as tendinous gap formation (mean: 0.7 mm), implying a preferable stability of the repair. Overall, this study demonstrates good ex vivo mechanical stability for a proposed suture anchor repair technique for type IV FDP tendon avulsion injuries, which might enable early postoperative mobilization in patients. The technique's subcutaneous implant placement and low implant load are expected to reduce potential complications observed in other commonly used repair techniques. This approach warrants further evaluation in vivo.

In 1977, Leddy and Packer1 introduced a classification system for flexor digitorum profundus (FDP) tendon avulsion injuries consisting of three avulsion types. Each of the types is defined by the level of FDP tendon retraction and potential concomitant bony injury1,2,3,4,5. Smith6 later added a fourth type to this classification describing a bony fragment avulsion associated with a tendon separation from the fragment and retraction along the flexor tendon sheath to a variable degree. Among tendon avulsion injuries in the hand, the probability of encountering a type IV avulsion of the FDP tendon is rather low and recommendations for surgical treatment are limited, substantiated mainly by case reports. Techniques and technique combinations ranging from distal interphalangeal (DIP) joint arthrodesis, K-wire fixation, interosseous sutures, screw fixation, pull-out wires to miniplate fixation techniques were proposed3,6,7,8,9,10,11,12,13,14. Consistent with postoperative treatment algorithms of flexor tendon injuries in general, the goal of a satisfying functional outcome can only be achieved by a rigid repair construct enabling a reasonable balance between protection of the repair and prevention of adhesion formation during early postoperative mobilisation15,16,17.

Due to the nature of this injury, repairs demand the combination of techniques attaining a rigid fixation of the avulsion fragment, as well as a solid reattachment of the retracted tendon1,18,19,20,21. Infections, repair construct disintegrations and secondary dislocations with loss of joint congruity leading to posttraumatic DIP joint arthrosis have been described primarily in external fixations. In contrast, internal fixation techniques have proven to be less prone to fulminant complications in other types of FDP tendon avulsions, seeming favourable for the treatment of a type IV injury as well. Suture anchor-based techniques were introduced to facilitate internal fixation of tendons and recent findings in solitary tendon- or bony FDP tendon avulsion repair seem promising10,18,22,23,24,25,26,27,28,29,30,31,32,33. Adopting findings from our previous study31 using a suture anchor combined with an interosseous suture pattern based on a tension banding principle to a type IV injury, we believe to attain a biomechanically stable repair avoiding reported complications.

The aim of this investigation was to conduct a biomechanical evaluation of a suture anchor technique, hypothesizing a sufficient rigidity and stability during early mobilisation simulation. In order to outline the advantages of this internal repair technique, we compared the biomechanical characteristics of the suture anchor technique to a double suture interosseous fixation and a screw repair combined with a single suture interosseous tendon fixation.

Although no specific biomechanical testing protocols for type IV FDP tendon avulsion were identified in the literature, a variety of procedures for injury simulation and biomechanical repair evaluation of other types of flexor tendon avulsions have been described30,31,34,35,36,37,38. Based on the methods described by our group in 201831 we embraced a curvilinear testing model to examine biomechanical characteristics during cyclic loading (postoperative mobilisation simulation) and when loaded to failure34,39. None of the tissues used in this study were procured from prisoners. All specimens were obtained at the Center for Anatomy and Cell Biology, Medical University of Vienna where all of the body donors provided written informed consent prior to their death for use in teaching and science. The Institutional Ethics Committee of the Medical University of Vienna approved this study. All methods were performed in accordance with the relevant guidelines and regulations.

Fifteen fresh-frozen hand-specimens (seven female, eight male) were randomly selected (mean 75, range 65–85 years of age). The specimens were thawed at 4 °C for 48 h prior to dissection. Intending to avoid bias originating from increased loss of bone density due to particular diseases donor records were reviewed. No signs of impaired bone quality were recognized during specimen preparation. From these 15 hand specimens, 45 fresh-frozen distal phalanges with an intact FDP tendon attachment from the ring, middle, and index fingers were exarticulated at the DIP joint and separated from surrounding soft tissues. The designated tendon length between the volar base of the distal phalanx and the point of tendon transection was 12 cm. At the level of the A4 pulley we determined the cross-sectional area (CSA) of the tendon measuring the tendon diameter and calculating the circle area in mm2. To allow full visualisation and further evaluation of the repair during cyclic testing, adjacent soft tissues and the A5 pulley were excised. Injury was simulated by separating the tendon from its bony insertion with a scalpel and creating a volar base avulsion fragment including at least 30% of the articular line using a chisel. Rehydration of the specimen was ensured by storage in saline solution (0.9% NaCl) until testing. Throughout preparation and testing, specimens were irrigated with this solution periodically.

Three study groups were formed by random allocation of 15 specimens to each group. In the first group (termed "anchor"), repair was conducted with a Corkscrew 2.2 × 4 mm suture anchor loaded with a #2-0 FiberWire (Arthrex, Naples, Florida) using the concept of an intratendineous and interosseous suture placement concordant with a novel repair technique presented in our previous study31 with the exception of addressing the complex injury characteristics of a type IV FDP tendon avulsion injury. This suture anchor technique follows a tension banding principle, increasing pressure on the avulsion fragment and the reattached tendon when strain is applied to the tendon and eventually the repair complex. The mean CSA of the FDP tendons was 6.1 ± 1.8 mm2. Intact cortical bone, approximately 1–2 mm distal to the bony avulsion site was considered the optimal anchor insertion area. A 45° retrograde insertion angle was chosen to reduce the probability of implant extrusion during loading of the repair29,36. After anchor placement, we guided each anchor suture through the distal stump of the tendon and subsequently inserted both sutures into the bony fragment at the borders between the median and the lateral thirds of its volar surface. Both sutures were then inserted into the corresponding site of the avulsion area of the distal phalanx. Following penetration of the dorsolateral cortex of the distal phalanx, sutures were retrieved on both sides of the distal phalanx. At this point, pulling on the sutures led to a reduction of the avulsion fragment and a reattachment of the tendon to its insertional footprint. We then inserted both sutures into the tendon in a transverse manner at the level of the volar base of the distal phalanx and tightly wove the sutures together along the median line of the FDP tendon. A Bunnell suture pattern was utilized to ensure a sufficient tendinous hold grasping approximately 1 cm of the distal tendon stump. A buried 5-throw square knot concluded the repair (Fig. 1).

(A) Specimen repaired with the suture anchor technique. Articular line alignment is visualized. (B) Illustration of the suture anchor technique depicting implant position, as well as the pattern of suture placement within the distal phalanx, the bony avulsion fragment and the tendon.

We reattached both injury components using two serial interosseous sutures in the second group (termed "interosseous"). The suture material, a #2-0 FiberWire suture (Arthrex, Naples, Florida), equalled the suture material used in the "anchor" group. The distal interosseous suture grasped the distal end of the FDP tendon stump, held the distal aspect of the avulsion fragment in its avulsion footprint, and was guided out through the intact dorsal corticalis of the distal phalanx. Main reduction of the articular line and compression of the fragment was achieved with the proximal interosseous suture. First, this suture was woven into the distal tendon stump using a Bunnell suture pattern. Subsequently the threads were guided through the avulsion fragment as well as into the intact portion of the distal phalanx. A 5-throw square knot at the dorsum of the distal phalanx proximal to the germinal matrix finalized each of the two sutures (Fig. 2). Calculation of the CSA in this group resulted in a mean of 7.1 ± 2.3 mm2.

(A) Illustration of the interosseous suture reattachment of the bony and tendinous injury components. To allow a clear perception of the suture placement within the tendon and the avulsed bony fragment in the top view, dorsal aspects of the sutures are not depicted. (B) Photographic image of a specimen repaired with interosseous sutures.

In the third group (termed "screw") we repaired the simulated injury combining a double screw fixation of the avulsed fragment and a single interosseous suture reattachment of the tendon. Two cannulated cortical titanium screws 2.0 × 8 mm (Arthrex, Naples, Florida) were inserted after verification of good fragment reduction and an aligned articular line. A #2-0 FiberWire suture (Arthrex, Naples, Florida) was woven into the proximal tendon stump using a Bunnell suture pattern. Approximately 5 mm distal to the level of screw insertion the threads were inserted through the reduced avulsion fragment into the intact distal phalanx approximating the tendon stump to the avulsion fragment. The suture was secured with a 5-throw square knot at the dorsal aspect of the distal phalanx sparing the germinal matrix (Fig. 3). The mean CSA of the tendons repaired with screws was 6.6 ± 1.4 mm2.

(A) Type IV injury repair using minifragment screws for bony avulsion repair and an additional interosseous suture for tendon reattachment. (B) Specimen after repair with screws and an interosseous suture. (C) Lateral fluoroscopy image of the distal phalanx after avulsion fragment repair with screws.

The biomechanical testing setup and the data acquisition protocol were adopted from our previous investigation31 evaluating repair techniques of type III FDP tendon avulsion injuries, seeming most suitable for the investigation of biomechanical characteristics of type IV injury repairs as well. To establish a firm hold of the specimens and to enable visualisation of the dynamics of the repair throughout biomechanical testing we used a distal phalanx (DP) holding cylinder mounted on an electromechanical tensile testing machine (Zwick Z050, ZwickRoell GmbH, Ulm, Germany) equipped with a 1 kN load cell. Prior to testing the load cell was checked for its accuracy in the low load region using calibrated weights. In the range from 0.1 to 5 N the maximum deviation was 7 mN. Following repair, each specimen was placed into the DP cylinder allowing full exposure of the repair through its semicircumferential window. Using common white plaster as a fixation material, all repair components remained unaffected, thus preventing unintended rotation or tilt of the specimen within the cylinder. A ball-joint linking the tensile testing machine to the cylinder ensured an optimal alignment of the specimen-repair construct along the longitudinal axis of the distal phalanx and the reattached tendon when strain was applied, preventing bias by translational forces to the repair site. Seven centimetres proximal to the repair, the tendon was retained using a standard tensile clamp. The articular line of the distal phalanx, of the avulsion fragment, as well as the tendon at the corresponding articular line level were marked with a blue felt tip pen under a baseline preload of 2 N allowing visual assessment of the repaired injury components (Fig. 4). Continuous cyclic loading from 2 to 15 N at a rate of 5 N/s for a total of 500 cycles simulated immediate postoperative mobilisation. Upon completion of cyclic loading, specimens were loaded to failure at a rate of 20 mm/min. A high-resolution camera placed in front of the experimental setup documented bony and tendinous gap formation at the repair site at the initial 2 N preload, periodically every 100 cycles, as well as, when loaded to failure, disclosing failure mechanisms. Consistent with methods used in our former studies30,31 on FDP tendon avulsion repair, the load at first noteworthy displacement (2 mm) representing the load at the occurrence of an avulsion fragment- or reattached tendon derived, clinically relevant global system displacement was evaluated. To allow a thorough perception of the repair- and failure biomechanics, following data were recorded and analysed as well: load at failure, elongation of the system, gap formation at the bone–bone contact line (proximalisation of the fragment resulting in an articular step off), gap formation at the bone–tendon insertion line, and the mechanism of failure. Reaching a structural balance of the tendon and the suture material after 50 cycles, elongation of the system was measured between the 50th and the 500th cycle. Bony or tendinous dynamics at the repair site documented on photographs were analysed with Image J Software (National Institutes of Health, Washington, DC, United States).

(A) Specimen preparation and injury simulation prior to repair. (B) Repaired specimen secured in the DP holding cylinder mounted on the tensile testing machine. Distance (r) indicating the 10 mm reference for further image evaluation.

Statistical analysis was performed using IBM SPSS Statistics Version 26, 64-bit. Normal distribution was assessed for every parameter by the analysis of histograms. Subsequently, for normally distributed data, one-way analysis of variance (ANOVA) was performed to detect any differences between the three repair techniques. Homogeneity of variance was given in every parameter; therefore, the Tukey post-hoc analysis for pairwise comparison was used. Analogously, for non-normally distributed data, the Kruskal–Wallis test and the Dunn´s post-hoc test were utilized. Data of normally distributed parameters are presented as mean and standard deviation, non-normally distributed data as median and interquartile range. p-values lower than 0.05 were considered statistically significant.

We were able to simulate the injury in each specimen in a consistent manner. When performing the repairs, no adverse events emerged resulting either from the repair procedure or from structural quality of the specimens. The biomechanical setup fulfilled our expectations on a rigid hold of the distal phalanx and a firm capture of the proximal tendon stump. Regarding the homogeneity in tendon size, we observed no statistically significant difference in CSA of the tendons between our study groups (p = 0.30). During cyclic loading no distinct failures occurred at the repair site such as extrusion of the distal phalanx out of the DP cylinder, premature implant tear-out, suture cut-out or tendon rupture, nor did we encounter any technical failure of the testing machine. In the course of each biomechanical testing, elongation was the first parameter to be evaluated. A displacement versus force curve was generated for each specimen enabling determination of elongation and potential displacement range throughout cyclic loading (Fig. 5). A statistically significant difference of elongation between all three study groups was detected in favour of the suture anchor repair technique (p = 0.001). The indicated median elongation in the "anchor" group was 0.79 (IQR 0.70–1.27) mm, 1.03 (IQR 0.74–1.27) mm in the "interosseous" group, and 1.6 (IQR 1.07–2.26) mm in the "screw" group, respectively. Photographic image analysis after 500 cycles revealed a mean gap formation at the bone-bone contact line of 0.1 ± 0.06 mm in the "anchor" group compared to 1.0 ± 0.41 mm in the "interosseous" group and 0.2 ± 0.22 mm in the "screw" group. The least bony gap formation was found in the "anchor" group (p =  < 0.001). Nevertheless, extent of bony gap formation between the "anchor" group and the "screw" group was not statistically significant (p = 0.580). At the bone-tendon insertion line, the least gap formation was observed using the suture anchor repair technique as well (p < 0.001). The measured bone-tendon gap was 0.68 ± 0.24 mm in the "anchor" group, 1.97 ± 0.57 mm in the "interosseous" group and 1.5 ± 0.61 mm in the "screw" group. By the time repairs were loaded to failure, the strain on the tendon increased and the load at first noteworthy displacement (2 mm) was captured on the displacement versus force curve. No statistically significant difference (p = 0.311) was observed between our three study groups (Table 1). While the strain reached a critical magnitude, we observed the failure to be a sequence of close events beginning with minimal negative spikes as a sign of suture material breakage and subsequently disintegration of the repair. Statistically significant higher load at failure values indicated the superiority of the suture anchor repair technique in load tolerance (p = 0.004). Repairs failed either by anchor pull-out from the bone or suture cut-out within the tendon when the suture anchor technique was applied. Interosseous suture repairs expectedly failed by suture cut-out in most cases, interestingly a fracture of the distal phalanx occurred as the mechanism of failure in 5 specimens. In the "screw" group, the failure mechanisms were either a screw pull-out or suture cut-out (Table 2).

Exemplary load versus displacement curve-graphs for each particular repair technique. The circle marks the point of measurement of the "load at the first noteworthy displacement (2 mm)". The triangle indicates the "load at failure" assessment point.

A considerable variety of repairs for type IV FDP tendon avulsions has been introduced so far. Primarily case reports on diagnostic approaches, as well as repair techniques in particular were identified in the literature7,8,9,10,11,12,13. No biomechanical studies on type IV injury repairs, nor clinical reports on suture anchor repairs of this injury were identified in the present literature. Reported repairs mainly based on K-wire or screw fixation of bony avulsion fragments and pull-out suture reattachment of the avulsed tendon. Each of these repair techniques holds a significant risk of complications and when combined, the incidence of complications may further increase1,18,19,20,32,34. Conclusions on therapy related outcomes are limited by omitted postoperative protocols and scarce consistency on outcome reports. Considering the findings of Leversedge et al.40 tendon nutrition may be impaired in a severely retracted tendon requiring timely repair. Therefore, we advocate additional imaging by ultrasound or MRI in each injury with a volar avulsion fragment of the distal phalanx.

The aim of this study was to introduce and biomechanically validate a firm and rigid repair technique using a single implant addressing both injury components; the bony fragment and the avulsed tendon. Moreover, an internal fixation technique is expected to result in a low incidence of peri- and postoperative complications, thus increasing the probability of a satisfying functional outcome. In recent literature, suture anchors moved into the focus of tendon repairs of the hand due to improved materials with a high load tolerance despite the small implant size18,22,23,30,34,41,42. A suture anchor based technique was already introduced for a type III injury repair31. The application appeared to be independent of avulsion fragment size, therefore supposedly preventing complications such as fragment breakage due to drilling or overtightening of screws. The superiority of the repairs’ biomechanical characteristics led us to the inquiry, weather it may be suitable for a type IV tendon avulsion as well.

The lower elongation in the "anchor" group reflects the formation of a solid conjunction of both injury components acting as a single avulsion element. This may imply a positive effect during postoperative mobilisation, as strain forces seem to be distributed equally between all affected structures. We divided the overall gap formation into two distinct parameters; gap formation at the bony avulsion site and at the bone-tendon insertion. Thus, we were able to draw conclusions on repair characteristics comparing the biomechanical data between study groups. Eglseder et al.8, as well as Trumble et al.13 used a single minifragment screw for a type IV injury repair. To improve fragment compression and ensure rotational stability, we decided to establish a double screw repair. Supporting our hypothesis of the suture anchor repair technique's rigidity, we did not detect a statistically significant difference of bony gap formation compared to the screw repair. The tension banding principle may be the primary cause for this finding. Considering the material properties of the sutures used and application of a comparable intratendinous suture pattern in all three study groups, we expected a similar gap formation at the bone-tendon insertion after 500 loading cycles. Interestingly, our observation revealed a superiority of the anchor technique. We believe the same sutures grasping the tendon twice and tightening it to the bone to be the most probable rationale. This parameter did not show any statistically significant difference between screw- and interosseous repairs, although the tendinous gap formation was the largest in the interosseous group. Highest load at the first noteworthy displacement was measured in the "interosseous" group. The suture anchor technique, as well as screw repairs seem to provide the most rigid hold in the lower loads region emphasizing their relevance during assisted postoperative mobilisation. Repair failure occurred at significantly higher loads in specimens treated with the suture anchor technique.

Concordant with our prior findings31 evaluating a suture anchor technique to repair a type III FDP tendon avulsion, we found the load at failure to be the highest and the gap formation to be the lowest when the same technique was applied. We previously observed a mean load of 100.5 N, whereas in this study a mean load of 72.7 N was detected when the anchor technique was loaded to failure. We believe that this difference originates from the injury composition, requiring the repair to stabilize not only a bony avulsion but a bony fragment and an avulsed tendon simultaneously. The tension banding principle of the suture pattern seems to be equally effective in preventing relevant bony gap formation in both injury types. In our prior investigation of type I injuries30, we observed mean failure loads of 65 N using an identical implant, securing the tendon to the bone applying a Bunnell suture pattern only. In general, due to increased resistance to tendon gliding resulting from postoperative soft tissue edema, intratendinous flexor tendon sutures are believed to require an initial strength of at least 30–55 N to allow gentle or moderate active finger motion43,44,45,46,47,48. The combination of the chosen implant, the suture material and the suture pattern in this study seems to outbalance the load tolerance of other distinct techniques addressing each injury component separately, underlining the biomechanical superiority of the here proposed technique.

Extracting the tendon out of the flexor tendon sheath and exarticulating the distal phalanx, as well as the age of the used specimens and their thawing are relevant limitations of our ex vivo study design. Screws with a diameter of 2 mm were the thinnest available. In clinical use, a screw diameter of 1.5 mm or 1.8 mm would be preferable, although we did not observe any avulsion fragment disintegration or distal phalanx fracture during repair or cyclic loading. Evaluation of load at first noteworthy displacement was based on a global system displacement measurement. To enhance data quality of this critical parameter, local displacement measurement would be preferable in forthcoming investigations. Assessment of repair dynamics under more physiologic circumstances, with the flexor tendon sheath preserved seems to be the most reasonable next step in future biomechanical investigations. To understand the tendency towards potential complications that cannot be anticipated in an ex vivo biomechanical study subsequent clinical trials will be essential.

In this first biomechanical study on type IV jersey finger repairs we validated the here presented testing setup and were able to underline the superior biomechanical characteristics of a suture anchor-based repair technique. Our findings indicate a potential benefit for the treatment of this challenging injury when applying our single implant solution due to a low probability of peri- and postoperative complications and sufficient load tolerance. However, future clinical investigations will disclose the suture anchor repair's behavior in vivo throughout the healing process, thus revealing its relevance in surgical practice.

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.

Leddy, J. P. & Packer, J. W. Avulsion of the profundus tendon insertion in athletes. J. Hand Surg. Am. 2, 66–69 (1977).

Article CAS PubMed Google Scholar

Leddy, J. P. Avulsions of the flexor digitorum profundus. Hand Clin. 1, 77–83 (1985).

Article ADS CAS PubMed Google Scholar

Robins, P. R. Avulsion of the insertion of the flexor digitorum profundus tendon associated with fracture of the distal phalanx. In A. A. O. S. Symposium on Tendon Surgery in the Hand 151–156 (1975).

Al-Qattan, M. M. Type 5 avulsion of the insertion of the flexor digitorum profundus tendon. J. Hand Surg. Br. 26, 427–431 (2001).

Article CAS PubMed Google Scholar

Stamos, B. D. & Leddy, J. P. Closed flexor tendon disruption in athletes. Hand Clin. 16, 359–365 (2000).

Article CAS PubMed Google Scholar

Smith, J. H. Jr. Avulsion of a profundus tendon with simultaneous intraarticular fracture of the distal phalanx–case report. J. Hand Surg. Am. 6, 600–601 (1981).

Article PubMed Google Scholar

Buscemi, M. J. & Page, B. J. Flexor digitorum profundus avulsions with associated distal phalanx fractures. A report of four cases and review of the literature. Am. J. Sports Med. 15, 366–370 (1987).

Article PubMed Google Scholar

Eglseder, W. A. & Russell, J. M. Type IV flexor digitorum profundus avulsion. J. Hand Surg. Am. 15, 735–739 (1990).

Article CAS PubMed Google Scholar

Ehlert, K. J., Gould, J. S. & Black, K. P. A simultaneous distal phalanx avulsion fracture with profundus tendon avulsion. A case report and review of the literature. Clin. Orthop. Relat. Res. 1992, 265–269 (1992).

Google Scholar

Kang, N., Pratt, A. & Burr, N. Miniplate fixation for avulsion injuries of the flexor digitorum profundus insertion. J. Hand Surg. Br. 28, 363–368 (2003).

Article CAS PubMed Google Scholar

Langa, V. & Posner, M. A. Unusual rupture of a flexor profundus tendon. J. Hand Surg. Am. 11, 227–229 (1986).

Article CAS PubMed Google Scholar

Takami, H., Takahashi, S., Ando, M. & Masuda, A. Flexor digitorum profundus avulsion with associated fracture of the distal phalanx. Arch. Orthop. Trauma Surg. 116, 504–506 (1997).

Article CAS PubMed Google Scholar

Trumble, T. E., Vedder, N. B. & Benirschke, S. K. Misleading fractures after profundus tendon avulsions: A report of six cases. J. Hand Surg. Am. 17, 902–906 (1992).

Article CAS PubMed Google Scholar

Henry, S. L., Katz, M. A. & Green, D. P. Type IV FDP avulsion: Lessons learned clinically and through review of the literature. Hand (N. Y.) 4, 357–361 (2009).

Article PubMed Google Scholar

Becker, H. Primary repair of flexor tendons in the hand without immobilisation-preliminary report. Hand 10, 37–47 (1978).

Article CAS PubMed Google Scholar

Becker, H., Orak, F. & Duponselle, E. Early active motion following a beveled technique of flexor tendon repair: Report on fifty cases. J. Hand Surg. Am. 4, 454–460 (1979).

Article CAS PubMed Google Scholar

Cullen, K. W., Tolhurst, P., Lang, D. & Page, R. E. Flexor tendon repair in zone 2 followed by controlled active mobilisation. J. Hand Surg. Br. 14, 392–395 (1989).

Article CAS PubMed Google Scholar

Skoff, H. D., Hecker, A. T., Hayes, W. C., Sebell-Sklar, R. & Straughn, N. Bone suture anchors in hand surgery. J. Hand Surg. Br. 20, 245–248 (1995).

Article CAS PubMed Google Scholar

Zook, E. G. Complications of the perionychium. Hand Clin. 2, 407–427 (1986).

Article CAS PubMed Google Scholar

Zook, E. G. & Russell, R. C. Reconstruction of a functional and esthetic nail. Hand Clin. 6, 59–68 (1990).

Article CAS PubMed Google Scholar

Kang, N., Marsh, D. & Dewar, D. The morbidity of the button-over-nail technique for zone 1 flexor tendon repairs. Should we still be using this technique?. J. Hand Surg. Eur. 33, 566–570 (2008).

Article CAS Google Scholar

Huq, S., George, S. & Boyce, D. E. The outcomes of zone 1 flexor tendon injuries treated using micro bone suture anchors. J. Hand Surg. Eur. 38, 973–978 (2013).

Article CAS Google Scholar

Huq, S., George, S. & Boyce, D. E. Zone 1 flexor tendon injuries: A review of the current treatment options for acute injuries. J. Plast. Reconstr. Aesthet. Surg. 66, 1023–1031 (2013).

Article CAS PubMed Google Scholar

Kapickis, M. New, "loop" suture for FDP zone I injuries. Tech. Hand Up Extrem. Surg. 13, 141–144 (2009).

Article PubMed Google Scholar

Schultz, R. O., Drake, D. B. & Morgan, R. F. A new technique for the treatment of flexor digitorum profundus tendon avulsion. Ann. Plast. Surg. 42, 46–48 (1999).

Article CAS PubMed Google Scholar

Sood, M. K. & Elliot, D. A new technique of attachment of flexor tendons to the distal phalanx without a button tie-over. J. Hand Surg. Br. 21, 629–632 (1996).

Article CAS PubMed Google Scholar

Teo, T. C., Dionyssiou, D., Armenio, A., Ng, D. & Skillman, J. Anatomical repair of zone 1 flexor tendon injuries. Plast. Reconstr. Surg. 123, 617–622 (2009).

Article CAS PubMed Google Scholar

Tripathi, A. K., Mee, S. N., Martin, D. L. & Katsarma, E. The, "transverse intraosseous loop technique" (TILT) to re-insert flexor tendons in zone 1. J. Hand Surg. Eur. 34, 85–89 (2009).

Article CAS Google Scholar

Bond, S., Rust, P. & Boland, M. The accommodation of bone anchors within the distal phalanx for repair of flexor digitorum profundus avulsions. J. Hand Surg. Am. 44(986), e981-986.e986 (2019).

Google Scholar

Halat, G. et al. Biomechanical characteristics of suture anchor implants for flexor digitorum profundus repair. J. Hand Surg. Am. 39, 256–261 (2014).

Article PubMed Google Scholar

Halat, G. et al. Introduction of a new repair technique in bony avulsion of the FDP tendon: A biomechanical study. Sci. Rep. 8, 9906 (2018).

Article ADS PubMed PubMed Central Google Scholar

McCallister, W. V., Ambrose, H. C., Katolik, L. I. & Trumble, T. E. Comparison of pullout button versus suture anchor for zone I flexor tendon repair. J. Hand Surg. Am. 31, 246–251 (2006).

Article PubMed Google Scholar

Tempelaere, C., Brun, M., Doursounian, L. & Feron, J. M. Traumatic avulsion of the flexor digitorum profundus tendon. Jersey finger, a 29 cases report. Hand Surg. Rehabil. 36, 368–372 (2017).

Article CAS PubMed Google Scholar

Brustein, M., Pellegrini, J., Choueka, J., Heminger, H. & Mass, D. Bone suture anchors versus the pullout button for repair of distal profundus tendon injuries: A comparison of strength in human cadaveric hands. J. Hand Surg. Am. 26, 489–496 (2001).

Article CAS PubMed Google Scholar

Manske, P. R. & Lesker, P. A. Avulsion of the ring finger flexor digitorum profundus tendon: An experimental study. Hand 10, 52–55 (1978).

Article CAS PubMed Google Scholar

Schreuder, F. B., Scougall, P. J., Puchert, E., Vizesi, F. & Walsh, W. R. The effect of mitek anchor insertion angle to attachment of FDP avulsion injuries. J. Hand Surg. Br. 31, 292–295 (2006).

Article CAS PubMed Google Scholar

Tanaka, H., Manske, P. R., Pruitt, D. L. & Larson, B. J. Effect of cyclic tension on lacerated flexor tendons in vitro. J. Hand Surg. Am. 20, 467–473 (1995).

Article CAS PubMed Google Scholar

Xu, L., Wang, M., Trenga, A., Grindel, S. & Daley, R. The ideal insertion site for the flexor digitorum profundus tendon in jersey finger repair: A biomechanical analysis. J. Hand Surg. Am. 2020, 15 (2020).

Google Scholar

Latendresse, K. et al. Cyclic testing of pullout sutures and micro-mitek suture anchors in flexor digitorum profundus tendon distal fixation. J. Hand Surg. Am. 30, 471–478 (2005).

Article CAS PubMed Google Scholar

Leversedge, F. J. et al. Vascular anatomy of the human flexor digitorum profundus tendon insertion. J. Hand Surg. Am. 27, 806–812 (2002).

Article PubMed Google Scholar

Brar, R. et al. Reattachment of flexor digitorum profundus avulsion: Biomechanical performance of 3 techniques. J. Hand Surg. Am. 39, 2214 (2014).

Article PubMed Google Scholar

Rehak, D. C., Sotereanos, D. G., Bowman, M. W. & Herndon, J. H. The Mitek bone anchor: Application to the hand, wrist and elbow. J. Hand Surg. Am. 19, 853–860 (1994).

Article CAS PubMed Google Scholar

Sapienza, A., Yoon, H. K., Karia, R. & Lee, S. K. Flexor tendon excursion and load during passive and active simulated motion: A cadaver study. J. Hand Surg. Eur. 38, 964–971 (2013).

Article CAS Google Scholar

McLarney, E., Hoffman, H. & Wolfe, S. W. Biomechanical analysis of the cruciate four-strand flexor tendon repair. J. Hand Surg. Am. 24, 295–301 (1999).

Article CAS PubMed Google Scholar

Moriya, T., Larson, M. C., Zhao, C., An, K. N. & Amadio, P. C. The effect of core suture flexor tendon repair techniques on gliding resistance during static cycle motion and load to failure: A human cadaver study. J. Hand Surg. Eur. 37, 316–322 (2012).

Article CAS Google Scholar

Robertson, G. A. & Al-Qattan, M. M. A biomechanical analysis of a new interlock suture technique for flexor tendon repair. J. Hand Surg. Br. 17, 92–93 (1992).

Article CAS PubMed Google Scholar

Schuind, F., Garcia-Elias, M., Cooney, W. P. 3rd. & An, K. N. Flexor tendon forces: In vivo measurements. J. Hand Surg. Am. 17, 291–298 (1992).

Article CAS PubMed Google Scholar

Silfverskiöld, K. L. & May, E. J. Gap formation after flexor tendon repair in zone II. Results with a new controlled motion programme. Scand. J. Plast. Reconstr. Surg. Hand Surg. 27, 263–268 (1993).

Article PubMed Google Scholar

Download references

This work was supported by Arthrex GmbH, Munich, Germany by providing implants used in all study groups. Study ID: EMEA19052.

Department of Orthopedics and Trauma Surgery, Division of Trauma Surgery, Medical University of Vienna, Vienna, Austria

Gabriel Halát, Lukas Leopold Negrin, Paul Lennart Hoppe & Stefan Hajdu

Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria

Ewald Unger

Institute of Materials Science and Technology, Faculty of Mechanical and Industrial Engineering, TU Wien, Vienna, Austria

Thomas Koch

Department of Anatomy, Center for Anatomy and Cell Biology, Medical University of Vienna, Vienna, Austria

Lena Hirtler

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

G.H. and S.H. researched the literature and conceived the study. G.H. wrote the main manuscript, performed the biomechanical testing and supervised the data analysis.L.N. and P.H. performed the statistical analysis of the data.P.H. analyzed the data acquired from the photographic images. T.K. supervised and performed the biomechanical testing and interpreted the data generated by the biomechanical testing machine. E.U. designed and produced the specimen holding cylinder. L.H. designed and supervised the specimen preparation. S.H. contributed to the study design development and manuscript composition.

Correspondence to Gabriel Halát.

The implants used in all study groups in this manuscript were provided by Arthex GmbH, Munich, Germany, following an Investigator - Initiated Research Agreement. As all used implants were provided by the same company, we believe that objectivity, integrity and the value of the results were preserved. No financial support, nor any other non - financial benefits were, or will be recieved. The authors declare no competing interest.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and Permissions

Halát, G., Negrin, L.L., Hoppe, P.L. et al. A suture anchor-based repair technique for type IV jersey finger injuries: a biomechanical investigation. Sci Rep 13, 3493 (2023). https://doi.org/10.1038/s41598-023-30373-w

Download citation

Received: 02 August 2022

Accepted: 21 February 2023

Published: 01 March 2023

DOI: https://doi.org/10.1038/s41598-023-30373-w

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.